Validation and process control for low energy electron beam irradiation
24 - 26 May, 2027
Irradiation with electron beams in the energy range lower than 300 keV is used in the polymer industry for curing and crosslinking. These beams are also used in the pharmaceutical and medical device industries for sterilization of surfaces. Both applications can require delivered doses to be traceable to national standards and that the irradiation process is validated.
The Risø course on “Validation and process control for low energy electron beam irradiation” focusses on radiation sterilization, and it provides the background for understanding the validation requirements of the international standards in particular the dosimetric aspects of:
- IQ/OQ, Installation / Operational Qualification
- PQ, Performance Qualification
- Routine monitoring and Process Control
The penetration of the electron beam at less than 300 keV is limited, and even thin dosimeters can interfere significantly with the beam to be measured.
These problems will be addressed during the course, and methods for establishing measurement traceability will be reviewed and tested.
24 - 26 May, 2027
Risø High Dose Reference Laboratory
DTU Health Tech
Technical University of Denmark, Risø Campus
DK 4000 Roskilde, Denmark
Download course REGISTRATION FORM
Download Full course INFO
Download course TIME SCHEDULE
Practical dosimetry exercises
Irradiation will be done using Risø HDRL’s low energy electron accelerator EbLab
During these exercises the participants will:
- Calibrate dosimeters to be used in the subsequent exercises,
- Carry out IQ / OQ measurements,
- Make PQ dose mapping studies in products,
- Make routine dosimetric process control.
The dosimeter systems that will be used during the course are:
- Radiochromic film dosimeters
- Alanine / EPR film dosimeters
Microbiological aspects
For radiation sterilization or decontamination, it is necessary to determine the required minimum dose for an effective process. Methods for selecting or substantiating a sterilization dose for medical devices are described in EN ISO 11137-2. These and other methods and their applicability at low energy electron irradiation will be reviewed and discussed.
Material effects
The effect of irradiation on polymers will be discussed. This forms the basis for determining a maximum acceptable dose for the irradiation process.
Lecturers:
Arne Miller, Risø High Dose Reference Laboratory
Mark Bailey, Risø High Dose Reference Laboratory
Frederik Steenstrup, Danish Technological Institute
For more information, see the Course Info document
Registration
Please fill in the registration form and send it to: armi@dtu.dk
Address:
Contact:
Arne Miller Specialist Consultant Department of Health Technology Phone: +45 46774224 Mobile: 2132 5613 armi@dtu.dk
Christina Ankjærgaard Head of Dosimetry, senior researcher Department of Health Technology Mobile: +45 61656929 cank@dtu.dk